Photometer
- Halogen lamp 6V 10W 1000 hour long
life
- 7 Interferential filters 1 optional position (340-405-450-500-546-578-630 nm.)
- Photoconduction - Si Photodiode UV enhanced
- Sample aspiration with peristaltic pump
- Auto cleaning system at start-up and shut-down
- Auto Zeroing system before each batch
- Auto centering filter system
- Filter wheel controlled by a stepper motor (0.2 mm. resolution)
- Automatic washing between samples
- Temperature PELTIER controlled (25° 30° 37°c) +/- 0,1°
- Quartz flow-cell 33 µl volume
|
Software
- Work-list menu (Demographic
- results - tests)
- Report by sample and by chemistry
- 240 chemistries in memory
- Help on line
- Easy to use
- Rapid access to the reading menu (only one key)
- Quality control on 3 levels
- Count-down internal watch up to 24 external incubation time
- Open system
- Self-test diagnostics
- Results on file
- Service operations for manual mode
- Up to 10 calibration points for non linear readings
- Archive of results and patient data (10 years stored in memory)
- Reaction curves
- Special mode for coagulation test with the I.N.R calculation
- Special mode for Elisa test with Gray Zone auto calculation and/or
logarithmic curve |
Hardware
- CPU ALI 6117 Dos compatible
- 2 Mb RAM
- Floppy disk 3.5" 1.44Mb
- VGA graphic display mono (9" CRT)
- Keyboard (82 keys)
- Parallel port
- Internal thermal printer (80C)
- Hard Disk 850 Mb or Superior
- Serial port RS232
|
Specification
- Abs - 0.020 to + 2.500 and 0.0002
resolution
- Drift 0.003 Abs/hour
- C.V. at 340 nm. 2.000 Abs better than 1%
- Sample aspiration from 300ml to 3000ml
set chemistry by chemistry
- A/D converter: 16 bit resolution |