Photometer
- Long life Halogen lamp 6V 10W 1000
hour
- 7 Interferential filters from 340nm to 70nm (band pass ± 2nm)
- Auto Zero and auto calibration for all the wavelengths
- Accuracy ± 1% from 0 to 3000 mAbs
- Linearity better than 0,5%
- Noise less than ±2 mAbs at 340 nm and 3000 mAbs |
Hydraulic
- High precision micro-processor
controlled Diluter module with a syringe of 1000 µl increasing 3 steps per
1 µl
- Accuracy ±0,5% from 3 µl to 1000 µl
- Self adjustment of mechanical tolerances
- 33ml Stanley steel flow
cell
- Aspiration by peristaltic pump
- Aspiration volume programmable by computer |
Sample
- 64 free positions, including 9
calibrator and/or control positions |
Reagents
- Till 18 reagent positions for each
run, 96 reaction positions (4 segments of 24 reaction tubes each) |
Temperature
- Cell temperature 37° C ±0,1° C
(PELTIER effect)
- Incubator temperature 37° C ±0,3° C (computer controlled) |
Chemistries
- 64 different tests available
- End-point, bi-chromatic, kinetic, fixed
time, differential, sample blank
- Calibration curve till 8 points for all the
tests. |
Hardware
- Built-in computer with self-test
- Firmware coded in EPROM no disk required
- On-board membrane keyboard
- VGA graphic display mono (9" CRT)
- Internal thermal printer (20C)
- PC DOS compatible computer external (option) and remote control till 20 meters
- Serial port RS232
|
Software
- Automatic self-test diagnostic
- Work-list available by internal or external computer
- Quality control for each test
- Batch optimization run or random run for STAT
- 120 test/hour (average of daily normal routine) |